GE Optima 650: cracked patient table drive belts
What happened
Both drive belts in the patient table were cracked and due for replacement, and the part number of one belt could not be matched from its marking.
What was checked
- The engineers read the markings on both belts under the table covers: one marking was usable, the other could not be matched to a catalogue number.
- They compared thickness, width and tooth type of the old belts and asked colleagues until both catalogue numbers were confirmed.
Cause
Not stated in the case. Ageing and wear of the rubber is the likely reason for the cracking, but this was not confirmed in the report.
What was done
The engineers replaced both belts. Reported in the case: the catalogue numbers were BELT 375 UP5M25 and BELT 350-2GT-3, and the advice was to identify an unmarked belt by its thickness, width and tooth type. The outcome was reported as solved.
Safety
Isolate the table from power before reaching into the drive: the table can move and the belts and pulleys are a pinch hazard, and run an electrical safety test before the table is returned to patient use.
Lesson: When a part marking cannot be read or matched, measure the old part — thickness, width and tooth profile — and confirm the catalogue number before ordering.