GE Optima: workstation fails to boot after BIOS reset

What happened

The scanner workstation would not load the system after the BIOS settings had been reset; the computer stalled at boot.

What was checked

Cause

The BIOS settings had reverted to defaults: the date and time were wrong and, as the engineers found in this case, the boot device priority was also off, so the computer could not find its boot device.

What was done

The engineers entered BIOS setup, set the time back and manually restored the boot device priority; in addition, the BIOS settings disc supplied with the scanner's original software set was used as a reference — every setting was compared against it and entered by hand. After that the system booted normally, and the case is reported as solved.

Safety

Before opening the workstation, shut the scanner down properly, isolate the computer from mains and UPS, and work with an anti-static strap; after restoring the BIOS, let the scanner complete its full self-test and warm-up before returning it to clinical use.

Lesson: A BIOS reset on a scanner workstation is rarely just the clock — restore the whole set of settings (time, boot order and everything else) against a documented reference before looking for other faults.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.