GE Saturn detector: error log points to the wrong module
What happened
In this case the log reported a fault in detector module 50, but the documentation said the real module number is half of the logged number, so it was unclear which module had actually failed and which part to prepare.
What was checked
- As reported in the case, the log pointed to module 50 while the documentation implied module 25, so the number itself could not be trusted on its own.
- The engineers swapped detector modules between positions to see whether the fault followed one module.
Cause
The engineers found that on older system versions the module number written into the error is twice the real module number, while on newer versions the logged number matches the actual module directly. This was observed in this one case with older software, not established as a rule.
What was done
The engineers localized the faulty module by swapping detector modules between positions: the error followed one specific module, which settled the question regardless of the logged number. The case was closed as solved.
Safety
Work on detector modules only with the system properly de-energised and gantry stored-energy precautions observed, and complete the manufacturer's prescribed calibration and electrical safety checks before the scanner returns to clinical use.
Lesson: A number in an error log is not always the physical module number — check how your software version maps logged numbers to hardware, or prove the fault by swapping before ordering any part.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.