GE Signa Creator 1.5: gradient error follows the control board between amplifiers
What happened
A gradient error kept recurring on a 1.5T MRI, and it moved with the gradient control board: whenever the engineers swapped the board between the three gradient amplifiers, the fault appeared in the amplifier where the board was installed.
What was checked
- Moved the suspect gradient control board in turn between the three gradient amplifiers; the error followed the board, which pointed away from the amplifiers themselves.
- Reviewed the recent ultracapacitor replacement in the gradient power supply and checked the supply's status LEDs, to rule out a power-side cause such as weak ultracapacitor support.
- The case discussion also covered protection trip thresholds; in the end no protection setting was changed — the fault turned out to be the board itself.
Cause
The engineers found a faulty gradient control board: an internal defect on the board triggered the fault indication regardless of which amplifier slot it occupied. The gradient power supply and its freshly replaced ultracapacitors were not at fault.
What was done
The gradient control board was replaced, and the scanner returned to normal operation. No protection settings were touched as part of the fix.
Safety
The gradient power supply stores lethal energy in large capacitors and ultracapacitors — isolate the system, allow the full discharge time from the service documentation, and verify absence of voltage before removing or installing any board.
Lesson: When a fault moves with a module instead of staying with a slot, the module is the culprit — swapping it between identical channels isolates it faster than chasing the subsystem where the error is displayed.