GE Signa TX HD: chiller coolant due for replacement

What happened

The chiller coolant on an MRI had reached the point of replacement, and the engineer needed the correct fluid composition for the refill.

What was checked

Cause

Reported in the case as routine coolant ageing: over time the additive package depletes and the fluid can support microbial growth and corrosion, which is why periodic replacement is called for. The exact interval was not discussed in the case.

What was done

The engineers replaced the coolant with the mixture reported in the case: ethylene glycol and water at about 35/65 by volume, with antimicrobial and corrosion-inhibitor additives. They advised confirming the exact grade and concentration against the manufacturer's service documentation for the specific installation before filling. The case was reported solved.

Safety

Isolate and lock out the chiller power supply and relieve loop pressure before opening the circuit; ethylene glycol is toxic, so wear gloves and eye protection and collect all drained fluid for disposal as chemical waste.

Lesson: Chiller coolant is a specified consumable: match the manufacturer's grade, concentration and additive package instead of filling with a generic antifreeze.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.