GE · Vivid 7

GE Vivid 7: dark, doubled image on the linear probe only

What happened

On a GE Vivid 7 ultrasound the image from the linear probe suddenly went dark and unreadable even at maximum gain, with stripes at the top and a doubled image, as reported in the case. Convex and phased array probes worked normally on the same system.

What was checked

Cause

Not finally confirmed. The engineers suspected a fault on the Relay board in the image processing path, most likely in the section serving the linear probe, after the probe, port, software reset and calibration had been ruled out.

The cause is probable and was not fully confirmed.

What was done

Repair or replacement of the Relay board was recommended, after checking the relevant resistors on the board. As reported in the case, this is a probable fix: the cause was never finally confirmed before the board decision.

Safety

Before opening the system for board-level work, isolate it from mains and let internal capacitors discharge, and after any board repair run a full electrical safety test, including leakage currents on the patient applied parts, before returning the system to use.

Lesson: When a fault follows one probe type but not the port, the software or calibration, suspect the shared board in that probe's processing path and confirm it before ordering parts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.