GE Vivid 7: won't start from the power button after a long idle
What happened
In the reported case, the scanner started normally from the power button after a short downtime but failed to start after being fully idle over a weekend. Once the engineers got it started with an interim workaround, it ran all day without faults.
What was checked
- The engineers checked the low-voltage rails (3 V and 5 V, as reported in the case) via the system's built-in diagnostics after the failed start.
- They compared behaviour after a long idle, when the reservoir capacitors fully discharge, against behaviour after a quick restart the same day.
- They noted that after the unit had been powered for a while (about an hour, as reported in the case), starting from the button returned to normal.
Cause
Probable, in this case: dried-out electrolytic capacitors in the power supply. After a long idle the capacitors fully discharge, and the 5 V enable voltage is missing or drops out, so the soft-start sequence never completes; on a quick restart the partly charged capacitors are enough for a normal start.
The cause is probable and was not fully confirmed.
What was done
The permanent fix is replacement of the aged electrolytic capacitors in the power supply — a procedure from the manufacturer's service documentation, done by authorised personnel. The interim measures used in the case (keeping the unit powered before the button press, and a forced start that bypassed the start/enable circuit) should be treated only as a stopgap and diagnostics, not a repair.
Safety
Treat the power supply as energised even after unplugging — mains electrolytic capacitors hold a dangerous charge, so verify discharge before touching anything, and run a full electrical safety check (protective earth and patient leakage currents) before the scanner returns to clinical use.
Lesson: A no-start that appears only after a long idle points to the power supply losing an enable rail on full discharge, so check the PSU rails first before blaming the main board — and never leave a start bypass in place as a workaround.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.