GE · Vivid 7

GE Vivid 7: won't start from the power button after a long idle

What happened

In the reported case, the scanner started normally from the power button after a short downtime but failed to start after being fully idle over a weekend. Once the engineers got it started with an interim workaround, it ran all day without faults.

What was checked

Cause

Probable, in this case: dried-out electrolytic capacitors in the power supply. After a long idle the capacitors fully discharge, and the 5 V enable voltage is missing or drops out, so the soft-start sequence never completes; on a quick restart the partly charged capacitors are enough for a normal start.

The cause is probable and was not fully confirmed.

What was done

The permanent fix is replacement of the aged electrolytic capacitors in the power supply — a procedure from the manufacturer's service documentation, done by authorised personnel. The interim measures used in the case (keeping the unit powered before the button press, and a forced start that bypassed the start/enable circuit) should be treated only as a stopgap and diagnostics, not a repair.

Safety

Treat the power supply as energised even after unplugging — mains electrolytic capacitors hold a dangerous charge, so verify discharge before touching anything, and run a full electrical safety check (protective earth and patient leakage currents) before the scanner returns to clinical use.

Lesson: A no-start that appears only after a long idle points to the power supply losing an enable rail on full discharge, so check the PSU rails first before blaming the main board — and never leave a start bypass in place as a workaround.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.