GE Vivid E9: red power indication, 3.3 V rail running high

What happened

The scanner showed a red power fault indication. The system logs, as reported in the case, recorded the 3.3 V rail running above its allowed window, up to 3.56 V.

What was checked

Cause

A failing power supply in the host computer whose 3.3 V output had drifted high. In this single case the overvoltage came from that supply, not from the main power subsystem or the host board itself.

What was done

The engineers replaced the computer's power supply with a second-hand unit. After the swap the red indication reportedly cleared, the voltage errors did not return, and the scanner went back into service. Note that a used supply has unknown history, so its outputs should be measured before it is trusted in service.

Safety

Isolate the scanner from the mains before opening the power compartment — PSU capacitors hold a charge after switch-off — and complete an electrical safety (earth leakage) test before returning the unit to clinical use.

Lesson: When a system-level monitor complains about one voltage rail, test the supply actually producing that rail first — in this architecture it was the host computer's PSU, not the main power board.