GE Vivid E9: ultrasound section does not power up
What happened
The main power supply does not bring up the ultrasound part of the system: indicator LEDs on the ultrasound boards stay dark, while the VER block keeps its normal indication.
What was checked
- Checked the indicator LEDs near the motherboard and on the VER block to see which part of the power-up chain is still alive.
- Visually inspected the ultrasound board closest to the power supply for bulged or burnt capacitors.
- Substituted the main power supply with a known-good unit and observed whether the system starts.
Cause
A fault in the main power supply (MPS): as reported in the case, the unit kept some standby indication alive but no longer delivered the rails needed to start the ultrasound section.
The cause is probable and was not fully confirmed.
What was done
The MPS power supply was replaced with a serviceable unit; the system started on the substitute supply, which supported the diagnosis, though the outcome was reported as probable rather than fully confirmed.
Safety
Before working in the power supply area, disconnect the system from mains and allow time for the bulk capacitors to discharge, and complete the electrical safety test before returning the scanner to clinical use.
Lesson: When some standby LEDs are lit but whole groups of boards stay dark, map the power-up chain by indicators first and confirm the supply by substitution before ordering parts.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.