GE Vivid E9: ultrasound section does not power up

What happened

The main power supply does not bring up the ultrasound part of the system: indicator LEDs on the ultrasound boards stay dark, while the VER block keeps its normal indication.

What was checked

Cause

A fault in the main power supply (MPS): as reported in the case, the unit kept some standby indication alive but no longer delivered the rails needed to start the ultrasound section.

The cause is probable and was not fully confirmed.

What was done

The MPS power supply was replaced with a serviceable unit; the system started on the substitute supply, which supported the diagnosis, though the outcome was reported as probable rather than fully confirmed.

Safety

Before working in the power supply area, disconnect the system from mains and allow time for the bulk capacitors to discharge, and complete the electrical safety test before returning the scanner to clinical use.

Lesson: When some standby LEDs are lit but whole groups of boards stay dark, map the power-up chain by indicators first and confirm the supply by substitution before ordering parts.