GE · Vivid E95

GE Vivid E95: cyclic band over the image on all probes

What happened

On a GE Vivid E95 ultrasound scanner, a cyclic band (striping) pattern overlays the image on every probe tried. No error codes were reported in the case, and wiggling the TGC sliders once made the picture react.

What was checked

Cause

Not confirmed. The working suspicion, based on this case only, is unstable TGC sliders on the front panel feeding jittering values into the image chain — similar to a case once seen on a Vivid E9. This remains a probable cause; nothing was proven.

The cause is probable and was not fully confirmed.

What was done

No repair was completed in this case. The agreed next step was to run the front-panel diagnostic utility built into the scanner software as described in the manufacturer's service documentation and watch the digital slider values; if the fault is confirmed, the TGC slider assembly would be replaced — work for authorised personnel.

Safety

Isolate the scanner from mains power before touching or disconnecting any front-panel connectors, and complete the manufacturer's electrical safety test before the unit returns to clinical use.

Lesson: When an artefact appears on all probes, suspect the shared part of the chain — including front-panel controls — and confirm the suspicion from the controls' own digital readouts before ordering parts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.