GE Vivid I: complete system freeze during operation
What happened
The system froze solid during normal operation and stopped responding. No error codes were reported. Suspecting the power supply, the engineers moved to measure it directly.
What was checked
- Measure the voltages on the power supply outputs and compare them against the values in the service documentation.
- Open the power supply and inspect the solder joints under a microscope, particularly around components that heat up in operation.
Cause
Cracked solder joints on components inside the power supply. As reported in the case, roughly 60% of the joints showed cracks — typical damage from years of thermal cycling and vibration, invisible without magnification.
What was done
The engineers removed the old solder and re-soldered the cracked joints inside the power supply; the job reportedly took about two days of bench work. According to the case, the machine has run without a repeat freeze for about two years since, and the same failure had been seen on this model before.
Safety
Disconnect mains and discharge the bulk and filter capacitors before touching anything inside a switched-mode power supply, and after any PSU repair run the electrical safety checks (protective earth and leakage currents) before the machine goes back to clinical use.
Lesson: A hard freeze with no error codes can start in the power supply — before blaming the main boards, open the PSU and inspect its solder joints under magnification.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.