GE Vivid S5/S6: no normal boot with damaged probe connectors

What happened

The unit would not boot normally and came up only in safe modes; the probe connectors were found cracked and mechanically damaged.

What was checked

Cause

Reported in the case: mechanical damage to the probe connectors — one had been crushed by an impact when the unit was transported with a probe still plugged in, and the remaining connectors had degraded gradually over time.

What was done

The engineers repaired and replaced the damaged connectors; after that the unit returned to normal operation, and the outcome was reported as solved.

Safety

Isolate the unit from the mains and follow the manufacturer's discharge procedure before removing any board, and because the probe connectors are part of the patient-applied circuit, complete the electrical safety (leakage current) checks before returning the unit to clinical use.

Lesson: Never transport a scanner with probes attached — an impact on a plugged-in probe can damage the connector badly enough to stop the whole unit, so unplug probes before moving the equipment and inspect the connectors at every service.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →