GE Vivid S6: no image on monitor after reassembly
What happened
As reported in the case: after a teardown and reassembly the unit powers up, the fans run and the 20 V, 12 V and 5 V rails are present, but the monitor shows nothing at all.
What was checked
- The engineers traced the whole video path reported in the case — from the graphics output connectors to the internal cables — and found no visible damage.
- They removed the memory modules, wiped the contacts and reseated them; one module had not been fully clicked into its slot.
Cause
A memory module was left half-seated after reassembly, so the system never started its video output.
What was done
The engineers reseated the memory modules firmly and the image returned immediately; no parts were replaced.
Safety
Disconnect the unit from the mains and wait for internal voltages to settle before opening the cabinet and touching boards or memory modules.
Lesson: After any disassembly, a dead video output is more often a half-seated module or cable than a failed board — reseat everything along the suspected path before chasing deeper faults.