GE · Vivid S6

GE Vivid S6: power-supply part number keeps disappearing

What happened

On a GE Vivid S6 the part number of the power supply kept dropping out. Rewriting the identification data in the supply's memory with an external programming tool helped only for about a week, as reported in the case, before the fault came back.

What was checked

Cause

Suspected cause: the non-volatile memory IC on the power supply was degrading and losing its stored part-number data; this was not confirmed in the case.

The cause is probable and was not fully confirmed.

What was done

Two options were proposed: updating the system software to a version that does not depend on the stored supply part number, or replacing the memory IC on the power supply. The case was still being resolved and the outcome was not confirmed.

Safety

Work inside a power supply only with the mains isolated and after the storage capacitors have discharged, and complete an electrical safety test (protective earth and leakage currents) before returning the unit to clinical use.

Lesson: If stored identification data keeps corrupting after repeated restores, suspect the memory chip itself and consider a software version that tolerates the missing data while you arrange the proper repair.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.