GE Voluson E8 BT13: second DVI/VGA monitor shows the full screen instead of the ultrasound area
What happened
On a second monitor connected via DVI/VGA the system displays the entire console screen, while only the ultrasound image area was needed; through S-Video the output shows only the scan area, as required.
What was checked
- The engineers searched the system menus for a setting that changes the output format of the second DVI/VGA monitor.
- They tried to limit the DVI/VGA output to the ultrasound area only.
Cause
The engineers found that on this software/platform version such a setting does not exist: DVI/VGA only mirrors the full console screen, and the scan-area-only output is tied to S-Video. This is a platform limitation, not a fault.
What was done
Workaround, reported as solved: the engineers selected a resolution/format for the second monitor matching the ultrasound area, so the duplicated screen fills the display with the scan image; for a strictly scan-area-only output, S-Video remains the option.
Safety
Before connecting or disconnecting video cables at the rear of the console, power the system down and isolate it from the mains.
Lesson: Not every display anomaly is a fault — some behaviour is fixed by the platform version, so check what the version actually supports and find a practical workaround before opening a defect case.