Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / GE Voluson E8
GE · Voluson E8

GE Voluson E8: herringbone interference on convex probe images

What happened

Sawtooth, herringbone-type interference appeared on images from both convex probes (the volume one and the regular one), while the vaginal probe produced a clean image. No error codes were reported.

What was checked

Cause

Not confirmed. The probable cause in this case was mains-borne interference from a nearby switching power supply, for example drivers of LED light fixtures, coupling into the system. The pattern — interference on the convex probes with a clean vaginal probe — argued against transducer failure, but the source was never finally proven.

The cause is probable and was not fully confirmed.

What was done

The recommendation in this case was diagnostic rather than a repair: power the unit from a UPS or on battery to confirm the noise enters via the mains, then locate and remove the interference source — move the unit to a clean circuit or deal with the offending device. The outcome was probable, not confirmed.

Safety

Before any mains-side checks, de-energise the unit and verify the protective earth path; never cure image noise by lifting or bypassing the ground conductor, and confirm the unit passes an electrical safety test before it returns to clinical use.

Lesson: Patterned interference limited to certain probes is more often picked-up mains noise than transducer failure, so test from a UPS or battery before ordering parts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes