GE · Voluson E8

GE Voluson E8: RFI board fault, FPGA reball needed

What happened

In this case the fault led to the RFI board of a Voluson E8, and the engineers' diagnosis came down to the FPGA on that board needing a reball; the discussion was whether to swap the board as a unit or repair it down to component level.

What was checked

Cause

The engineers' diagnosis: defective solder joints under the FPGA — the large BGA device in the centre of the RFI board — a typical board-level fault where thermal cycling and mechanical stress crack the solder balls under the chip.

The cause is probable and was not fully confirmed.

What was done

The proposed fix was a reball of the FPGA: the chip is lifted off the RFI board, its solder balls are renewed and the chip is refitted — component-level repair instead of replacing the whole board. The outcome was reported as probable; the case did not confirm a finished, verified repair, and such rework needs proper BGA equipment and skills.

Safety

Isolate the scanner from the mains and let internal capacitors discharge before opening it, work on the board under ESD control, and repeat the electrical safety tests (protective earth continuity and leakage currents) before the machine returns to clinical use.

Lesson: A component-level repair such as an FPGA reball can save an expensive board, but only when the fault is confirmed on that board and the rework is done with proper BGA tools and verified before return to use.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.