GE Voluson: no proper boot after BIOS settings reset
What happened
On a GE Voluson the BIOS settings kept resetting, and after the fault the unit would not start up correctly, including when running on the internal battery.
What was checked
- Confirm that BIOS settings are not retained after power cycles and that startup on battery power fails after the reset.
- Use the BIOS recovery / restore-optimal-defaults routine as described in the manufacturer's service documentation, then verify that settings persist after a reboot.
Cause
The engineers found a reset/corruption of the stored BIOS settings: with the stored configuration lost, the boot parameters were invalid, which in this case also blocked correct startup on battery power.
What was done
The fix was to restore the manufacturer's default BIOS settings through the built-in recovery option — a procedure from the manufacturer's service documentation, done by authorised personnel. Reported in the case: after the restore the system started normally again, including on battery power, and the case was closed as solved.
Safety
Before any work inside the system unit, disconnect the machine from mains and the internal battery and observe ESD precautions, and run an electrical safety test before returning the scanner to clinical use.
Lesson: A BIOS settings reset can mimic a hardware failure — before swapping parts, check the stored setup and restore it from the service documentation.