GE Voluson S8: intermittent right trackball button
What happened
On a Voluson S8 the right trackball button worked with dropouts, cutting out periodically during use; the engineers also expected the trackball itself to degrade soon, and noted similarly low-grade control-panel components on a newer unit of the same model.
What was checked
- The engineers asked the manufacturer which control-panel spare parts are available and in what form — as piece parts or only as assemblies.
Cause
Wear with an intermittent break in the trackball button; in this single case the engineers attributed the early failure to low component quality of the panel.
What was done
Solved by replacing parts individually: the manufacturer offers the panel board, the silicone button membrane, removable buttons and encoders as separate items, so only the worn part is needed — far cheaper than the full keyboard assembly (reported in the case at over 2000 USD, with the membrane recently bought for about 600 USD, and prices and lead times reported as high overall). The button fault was resolved; further trackball wear remains likely.
Safety
Before removing the control panel, power the system down and disconnect it from the mains, use ESD grounding when handling the exposed boards, and run the electrical safety test before returning the unit to service.
Lesson: Before ordering an expensive control-panel assembly, ask the manufacturer for the piece-part list — buttons, membranes and encoders are often available separately at a fraction of the assembly price.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.