GE Voyager / Optima 450: grainy DWI as the diaphragm drifts during breath-hold

What happened

On breath-hold DWI the diaphragm dome moved between the upper and lower edges of the navigator window because breath-hold depth varied from hold to hold, and the images showed slight grainy noise; navigator-based respiratory synchronisation was unavailable or would not enable.

What was checked

Cause

Varying depth of the patient's breaths; the engineers' working assumption — unconfirmed in this case — was that the navigator could not keep up with these small diaphragm displacements.

What was done

The engineers found that the manufacturer's standard axial breath-hold DWI library protocol with a breath-hold of up to 20 s and increased averaging worked in this case; parameters reported in the case were NEX 2 at b50 and NEX 11 at b1000, with a scan time of about 4 minutes. This option was judged suitable for patients who cannot breathe rhythmically for triggering, such as patients with a stoma.

Safety

With breath-holds of up to 20 s, confirm throughout the sequence that the patient can hear instructions and reach the emergency call buzzer, and stop the scan if the patient cannot tolerate the hold.

Lesson: When respiratory navigation cannot lock onto irregular breathing, a breath-hold protocol with extra averaging is usually the more robust answer than fighting the trigger.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.