Genoray C-arm: factory weak points on a new unit

What happened

No single fault here: on a new C-arm the engineers found a set of factory weak points showing up in normal use — a detector connector not fully seated, blood running under the tube overlay, brittle brake-handle bolts, and thin plastic tube covers.

What was checked

Cause

Manufacturer design shortcomings and weak stock components: not one defect but a set of factory weak points that surface one by one during normal operation, as the engineers found in this case.

What was done

Preventive rework at commissioning: thread locker on all threaded joints, replacing the detector's LAN connector with a switch fed from the motor power supply via a DC/DC converter, a silicone gasket under the tube overlay to stop blood ingress, higher-strength brake-handle bolts, protective locks on the tube covers, and a software backup before starting. The unit was then used without these issues; the outcome was reported as solved.

Safety

Blood under covers is a biohazard — clean and disinfect wearing gloves before disassembly — and after any change to covers or wiring on an X-ray unit, verify earth continuity, leakage currents and correct refit of all protective parts with an electrical safety test before the system returns to clinical use.

Lesson: A brand-new unit can still hide factory weak points — walk through them and do preventive rework at commissioning instead of waiting for each one to fail in the field.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.