Genoray C-arm: factory weak points on a new unit
What happened
No single fault here: on a new C-arm the engineers found a set of factory weak points showing up in normal use — a detector connector not fully seated, blood running under the tube overlay, brittle brake-handle bolts, and thin plastic tube covers.
What was checked
- Check the seating of every connector at handover — the detector link was found not fully seated.
- Inspect threaded fasteners for looseness and secure them with thread locker.
- Open and inspect the tube-head overlay for fluid paths — blood had been running under it.
- Check brake-handle bolts for brittleness and swap them for a higher strength class.
- Assess the thin tube-cover plastic and fit protective locks (four covers in this case, as reported).
- Review the detector network link — the plain LAN connector was replaced with a switch powered from the motor supply through a DC/DC converter (28 V, as reported in the case).
- Back up the software before any rework.
Cause
Manufacturer design shortcomings and weak stock components: not one defect but a set of factory weak points that surface one by one during normal operation, as the engineers found in this case.
What was done
Preventive rework at commissioning: thread locker on all threaded joints, replacing the detector's LAN connector with a switch fed from the motor power supply via a DC/DC converter, a silicone gasket under the tube overlay to stop blood ingress, higher-strength brake-handle bolts, protective locks on the tube covers, and a software backup before starting. The unit was then used without these issues; the outcome was reported as solved.
Safety
Blood under covers is a biohazard — clean and disinfect wearing gloves before disassembly — and after any change to covers or wiring on an X-ray unit, verify earth continuity, leakage currents and correct refit of all protective parts with an electrical safety test before the system returns to clinical use.
Lesson: A brand-new unit can still hide factory weak points — walk through them and do preventive rework at commissioning instead of waiting for each one to fail in the field.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.