Genrui · KT-6300

Genrui KT-6300: WBC/RBC clog alerts and low counts after electrode wire repair

What happened

The needle washer block cracked and washer fluid dripped onto the WBC/RBC electrode wiring, which corroded through. After the wire was repaired, the analyzer kept flagging WBC and RBC clogs and produced very low results.

What was checked

Cause

The engineers found that the reagents had spoiled from improper storage while still inside their printed shelf life; per the case, opened reagent is reportedly good for only two to three months. Degraded reagent produced clog-type alerts and low counts that mimicked an obstruction.

What was done

With the wire proven good and the sample path visually confirmed, the engineers replaced the reagents with fresh stock; the clog alerts and low counts cleared and the analyzer returned to normal operation. Case solved.

Safety

Treat any fluid that leaked from the washer block — and everything it touched, including the electrode wiring — as potentially biohazardous: decontaminate, dry and inspect the area for corrosion, and complete electrical safety testing before returning the analyzer to use.

Lesson: When clog alerts and low counts persist after a fluid-path repair is proven sound, try fresh reagents before tearing into the hydraulics again — a spoiled reagent can look exactly like an obstruction.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.