Hamilton C2: New original sensor fails power-on self-test
What happened
On a unit reported as an older build, a newly fitted original sensor fails the startup self-test and triggers an error, even though the part is genuine and correct for the model.
What was checked
- Confirm the replacement sensor is original and identify its hardware revision.
- Check the compatibility of the sensor revision with the firmware version installed on the unit — reported in the case as the deciding check.
Cause
The engineers found that the unit runs older firmware intended for revision 1 sensors, while newly supplied original sensors are made as revision 2, so the startup test rejects the new part.
The cause is probable and was not fully confirmed.
What was done
The probable fix, as reported in the case, is updating the unit's firmware so it supports revision 2 sensors — a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Do not return the ventilator to clinical use until the full power-on self-test passes with the new sensor, ventilation is verified on a test lung, and the electrical safety test before return to service is completed.
Lesson: Before condemning a new sensor that fails a self-test on older equipment, check its hardware revision against the installed firmware — the part may be correct but not supported.