Hamilton · C2

Hamilton C2: New original sensor fails power-on self-test

What happened

On a unit reported as an older build, a newly fitted original sensor fails the startup self-test and triggers an error, even though the part is genuine and correct for the model.

What was checked

Cause

The engineers found that the unit runs older firmware intended for revision 1 sensors, while newly supplied original sensors are made as revision 2, so the startup test rejects the new part.

The cause is probable and was not fully confirmed.

What was done

The probable fix, as reported in the case, is updating the unit's firmware so it supports revision 2 sensors — a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Do not return the ventilator to clinical use until the full power-on self-test passes with the new sensor, ventilation is verified on a test lung, and the electrical safety test before return to service is completed.

Lesson: Before condemning a new sensor that fails a self-test on older equipment, check its hardware revision against the installed firmware — the part may be correct but not supported.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.