Hamilton · C3

Hamilton C3: real-time clock failure after long idle

What happened

A Hamilton C3 ventilator that had stood idle for a long time reported a 'Realtime clock failure' at start-up, as described in the case.

What was checked

Cause

The engineers found that the battery backing the real-time clock had discharged during the long idle period, so the clock lost its settings and the unit raised the fault at start-up.

The cause is probable and was not fully confirmed.

What was done

The reported fix was replacement of the clock backup battery (part number 369106, as reported in the case). A software update to version 2.0.8, as noted in the case, was also planned. The outcome was described as probable: the engineers expected the fault to clear after the battery swap, but this was one case, not a guaranteed rule.

Safety

Disconnect the ventilator from mains power and from the patient before opening the unit, and after the battery replacement complete the electrical safety test and the manufacturer's operational verification before returning it to clinical use.

Lesson: After a long storage period, check the real-time clock backup battery on a ventilator first, before chasing deeper faults.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.