Hamilton G5: "battery discharged" message after battery replacement
What happened
Reported in the case: after a battery was replaced on a Hamilton G5 ventilator, the unit kept indicating the battery as discharged. On a second identical unit, a new battery was accepted immediately.
What was checked
- The engineers found that in the reported case the battery had been swapped with the unit de-energised (a "cold" swap).
- The suspect battery was compared against a second identical G5, where it was recognised and charged normally.
Cause
Probable cause, not finally confirmed: the fault indication was latched because the battery was replaced with the unit powered down, instead of the manufacturer's documented replacement procedure with AC mains connected.
The cause is probable and was not fully confirmed.
What was done
As reported in the case, the recommendation is to change batteries per the manufacturer's documentation, with the unit on AC mains and batteries swapped one at a time. Where the fault persisted, the unit was handed over to the authorised service organisation. The resolution in this case was probable, not fully confirmed.
Safety
Before replacing batteries on a ventilator, confirm the unit is on AC mains and the remaining battery has enough charge to carry the patient through the swap; if that cannot be guaranteed, move the patient to a backup ventilator first.
Lesson: A "battery discharged" alarm appearing straight after a battery change often points to how the swap was performed, so verify the battery on a known-good unit and the replacement procedure before condemning parts.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.