Hamilton · G5

Hamilton G5: water in the air inlet line

What happened

The air inlet of a Hamilton G5 ventilator was found flooded with water, and the moisture was traced back to the upstream air supply line.

What was checked

Cause

Reported cause in this single case: the upstream air compressor was pushing water into the supply line when worn or not serviced on time; the conclusion was judged probable, not confirmed by measurements.

The cause is probable and was not fully confirmed.

What was done

A water trap (moisture separator) was fitted on the inlet line, and the compressor was put on disciplined scheduled service. The outcome in the reported case was considered probable.

Safety

After any liquid ingress keep the ventilator out of service until it is completely dry, passes the electrical safety test and a full pre-use self-test, and the gas path has been cleaned per the instructions for use.

Lesson: Guard every ventilator gas inlet against upstream moisture: a water trap plus disciplined compressor maintenance is far cheaper than water inside the machine.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.