HemoCue (микроанализатор) · MicroCC 18

HemoCue MicroCC 18: valve-clog alarms at startup and a leak at the RBC chamber

What happened

On startup the analyzer reported 'valve clog' alarms on lines XB3, XB4 and XB5 (as reported in the case). Replacing the tubes and checking the valves changed nothing, and during inspection the engineers also found a leak at the RBC measurement chamber, close to the electrode.

What was checked

Cause

In this single case the root cause was traced to the power supply, not to the tubes or valves: once it was sorted out, the startup 'clog' alarms on XB3–XB5 reportedly went away. The engineers treated the leak at the chamber electrode as a separate finding, not the root cause of the alarms.

What was done

The power supply was replaced or repaired (as reported in the case), after which the analyzer started without the valve-clog alarms and returned to service. Any remaining chamber leak should be corrected by a proper repair or chamber replacement per the manufacturer's service documentation — not by sealing it with epoxy on site.

Safety

Treat every part of the fluid path and any leaked sample as potentially infectious; and since fluid was found near electronics with a faulty power supply in the story, complete the electrical safety check, including leakage current, before the analyzer goes back into use.

Lesson: A 'valve clog' alarm is not always a fluid problem — verify what actually feeds the valves before swapping tubes, and a leaking measuring chamber needs a proper repair, not a dab of sealant.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.