Hitachi Airis Mate 0.2: low SNR on images

What happened

The scanner produced images with noticeably low SNR and no error codes. The engineers suspected the RF path and looked at the excitation side and the resonance curve first, to rule out a fault in the transmit chain before blaming the room environment.

What was checked

Cause

In this case the noise came from outside the scanner: switching-mode power supplies of the LED panel lights radiated broadband interference that reached the receive path and depressed SNR. The engineers also mentioned the temperature in the technical room as a possible additional factor, which was not confirmed.

What was done

The interference source was found and eliminated: the LED lighting power supplies were removed/replaced so that their switching noise no longer reached the RF path, and SNR returned to normal, as reported in the case. Any changes to equipment in the MR environment were handled according to the site's MR safety rules.

Safety

Only equipment verified as MR-compatible may be brought into the magnet room — a mains-powered instrument or luminaire can become a projectile or an interference source — and the system must pass its electrical safety checks before return to use.

Lesson: Low SNR on an open MRI often points outside the machine, so survey the room's electrical environment — especially switch-mode power supplies — before overhauling the RF chain.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.