Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Hitachi Arietta 65 / Arietta Vetus
Hitachi · Arietta 65 / Arietta Vetus

Hitachi Arietta 65 / Arietta Vetus: beamformer chip thermal failures

What happened

On an ultrasound unit of this model line, the beamformer (the beam-processing board) failed with signs of thermal damage to its chips. As reported in the case, most units of one batch showed the same fault, and the manufacturer ran a recall campaign.

What was checked

Cause

Probable cause: thermal failure (thermal cycling with loss of solder contact) of the beamformer chips. The recall mentioned in the case points to a systematic batch-level problem rather than random wear, but the exact mechanism was not confirmed.

The cause is probable and was not fully confirmed.

What was done

No repair solution was given in the case. As reported, the fault was handled through a manufacturer recall campaign. Any rework of such a board is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Before opening the cabinet, disconnect the unit from mains and allow the power-supply capacitors to discharge; handle the beamformer board with ESD precautions, and complete an electrical safety test before the unit returns to use.

Lesson: When the same failure appears on most units of one batch, treat it as a systematic fault and escalate to the manufacturer instead of chasing one-off board repairs.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes