Hitachi Avius: blue screen on boot, graphics driver crash
What happened
On normal boot the ultrasound system dropped to a blue screen with the crash pointing at the graphics driver. In safe mode with the basic VGA driver the unit booted and ran normally. In this unit the graphics subsystem is integrated into the motherboard, not a separate card.
What was checked
- Cleaned the contacts of the graphics module and reseated it, as reported in the case.
- Tried the suspect graphics hardware in a conventional PC — of limited value here, since the graphics is soldered onto the motherboard.
- Safe mode with the generic VGA driver confirmed the rest of the system ran normally, steering suspicion toward the graphics hardware rather than Windows or the application.
Cause
Not confirmed. The engineers considered failing video memory of the integrated graphics the most likely trigger of the driver crash; the final root cause was not established in this case.
The cause is probable and was not fully confirmed.
What was done
No completed repair is recorded in the case. Engineers who met a similar fault reported that replacing the graphics hardware resolved it; here, with the graphics integrated into the motherboard, that means a board-level repair or motherboard replacement following the manufacturer's service documentation, done by authorised personnel. The outcome was therefore probable, not confirmed.
Safety
Disconnect the unit from mains and observe ESD precautions before touching internal boards, and before returning it to service run the manufacturer's electrical safety checks (protective earth continuity and leakage currents) together with a functional image test.
Lesson: Safe mode with a generic VGA driver quickly separates driver trouble from failing graphics hardware, but confirm what hardware is actually fitted before ordering a replacement.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →