Hitachi Scenaria: very slow image transfer to PACS
What happened
A CT study of about a thousand images (as reported) went to PACS in batches of three images at a time, so the transfer dragged on; from another workstation the same PACS received data in seconds. The network was believed to be gigabit and the DICOM settings (AeTitle, ports) were correct.
What was checked
- Confirm the receiving PACS and whether scanner and PACS sit in the same subnet; compare against another workstation sending to the same PACS.
- Review image types and transfer syntax: whether compression is applied on send and whether RAW data is being transmitted.
- Look for TCP no-delay and batch transfer options in the send configuration (not found in this case).
- Rule out antivirus software interfering with the sending host.
- Send from the scanner and from the other workstation at the same time: when both slowed down, the shared network link became the suspect.
- Check the actual negotiated link speed at the wall outlet and switch port, not the labels on them: the link was running at 100 Mbit instead of gigabit.
Cause
The bottleneck was the physical network link: the wall outlet/port in the scanner room negotiated only 100 Mbit instead of gigabit, so every DICOM association crawled. All settings above the cable were correct — the engineers found the fault at the physical layer.
The cause is probable and was not fully confirmed.
What was done
The link was brought back to gigabit by reworking the defective wall outlet/patch connection, after which the slow send disappeared; no DICOM setting on the scanner was changed. The outcome was reported as probable, so verify the transfer speed after the repair.
Safety
Tracing cabling behind a CT means working near mains-powered equipment: handle only the low-voltage data leads and patch ports, keep clear of power outlets and scanner connections, and if any step requires entering or opening the scanner, leave it to authorised personnel and perform the electrical safety test before return to use.
Lesson: When one device sends to PACS slowly while another reaches the same PACS quickly, check the negotiated link speed at the wall port and switch before touching any DICOM settings.