Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Hologic Selenia
Hologic · Selenia

Hologic Selenia: workstation failure after full-disk imaging attempt

What happened

After an attempt to make a sector-by-sector image of the hard drive, the Sun/Solaris workstation stopped starting normally. NVRAM seemed to have reset to defaults and an incorrect MAC address in OpenBoot was suspected.

What was checked

Cause

The engineers found that the MAC address and host ID were actually normal. The real cause was the sector-by-sector copy and mounting of the whole disk: as reported in the case, partitions 1 and 2 on these Sun systems are service partitions, and only partitions 0, 3, 4 and 5 are meant to be worked with, so handling the entire disk broke the system configuration.

What was done

Do not mount or image the disk as a whole; work only with the usable partitions (0, 3, 4 and 5 in this case). Once only the correct partitions were used, the system was started and was reported running normally.

Safety

The workstation drives X-ray exposure, so before the mammograph returns to clinical use confirm that the exposure interlocks and emergency stop still work, and complete the electrical safety test and functional verification after any work on the workstation.

Lesson: On Sun/Solaris workstations never image or mount the whole disk: clone and restore only the data partitions, because service partitions are part of the system layout.

Covered in the course: Mammography Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes