Horiba Micros 60: high platelet background with low RBC counts
What happened
On a Horiba Micros 60 hematology analyzer the platelet background was reported in the case as high as about a hundred, counting was unstable, and RBC came out roughly halved when a sample was run.
What was checked
- Clean the large piston, its cup and the piston seal, then regrease the seal.
- Run a series of control measurements — in the case the engineers suggested watching stability over 20–40 runs; if results wander, plan to replace the seal.
- Open the dilution chamber and inspect the aperture under a microscope for build-up or cracks.
- Inspect the electrode wire and its soldered joint for corrosion or a broken connection.
Cause
Not finally established in this case. The likely suspects were a dirty or worn piston seal in the RBC/PLT dilution chamber, a partially blocked or damaged aperture, or a corroded solder joint on the electrode wire — any of these can raise platelet background and pull RBC counts down. The engineers treated it as one case, not a rule.
The cause is probable and was not fully confirmed.
What was done
Recommended in the case: clean the piston, cup and seal and regrease the seal; if counting stays unstable on the control runs, replace the seal; inspect the aperture under magnification and clean or replace it if coated or cracked; check the electrode wire and repair the solder joint if corroded. The outcome was reported as probable — a working path from one case, not a confirmed root cause.
Safety
Treat every surface wetted by samples as potentially infectious — wear gloves and eye protection while the chamber is open — and unplug the analyzer before touching the electrode wiring or soldering.
Lesson: When platelet background and depressed RBC appear together on an aperture counter, suspect the dilution chamber's piston, seal and aperture first, service them as a set, and confirm the repair with a run of control measurements.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.