Huvitz HRK-7000: power supply degraded by dried-out electrolytic capacitors
What happened
The power supply of the autorefractor-keratometer had degraded: its electrolytic capacitors had dried out and their ESR had risen far above normal.
What was checked
- Measure the ESR of the power-supply electrolytics with an ESR meter: reported in the case, the 1000 µF capacitors read around 0.7–0.8 Ω and the 33 µF ones had dried out completely.
- Judge each capacitor by its measured ESR, not by appearance — dried-out electrolytics often show no bulging or leakage.
Cause
In this case the engineers found aged, low-grade electrolytic capacitors in the stock power supply: over time the electrolyte had dried out, ESR had climbed, and the supply could no longer hold stable output.
What was done
Most of the power-supply electrolytic capacitors were replaced with quality low-ESR or solid-polymer parts of the same capacitance and voltage rating; the repair solved the problem.
Safety
Before working on the supply, unplug the instrument, let the electrolytics discharge and confirm with a meter that no voltage remains, then run the electrical safety test (earth bond and leakage current) before returning the instrument to use.
Lesson: When an older instrument misbehaves, test the power-supply electrolytics with an ESR meter early — capacitors that look fine but have dried out are a common hidden root cause.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.