Incart KT-04-3 Holter: MMC card goes write-protected, ECG recordings cut short
What happened
As reported in the case, the recorder captured only about 20 minutes of ECG instead of a full study, and the MMC flash card then refused to format, reporting the disk as write-protected. Attempts to clear the protection through standard PC tools had no effect.
What was checked
- Tried clearing the write-protect flag with standard PC tools (registry, diskpart) — the card stayed locked.
- Tried formatting from the Holter's own software — the same write-protection message.
- Tested a new card in the recorder to confirm the fault sat with the old card, not the device.
Cause
The engineers found the MMC card itself was dying: when flash memory wears out, the card's controller can latch the card into a permanent write-protected state, so the lock is a symptom of degradation, not a setting. In this case the recorder was healthy.
What was done
Replace the MMC card with a new one. Format the new card from the Holter software first; a PC format to FAT32 also worked in this case without problems, but the device-side format is the safer default. The recorder went back into service — case closed as solved.
Safety
Power the recorder down and disconnect it from the patient before swapping the card, and complete the manufacturer's electrical safety check for patient-applied parts before the unit returns to use.
Lesson: A removable card that suddenly goes write-protected mid-study is usually the card failing — replace it instead of fighting the lock.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.