Инкарт · Кардиотехника-07-3/12

Inkart Kardiotechnika-07-3/12: new ECG lead cables show as 'unknown'

What happened

After fitting new Holter lead cables, the recorder listed them as 'unknown' in the settings and did not recognise them at all.

What was checked

Cause

Not a cable defect. As the engineers found in this case, the new leadsets are only recognised on the current software/firmware version, and the recorder was running an older build.

What was done

A full software update to the latest build solved the problem, as reported in the case: the update installed cleanly on Windows 10, while on Windows 7 a full system upgrade to the latest builds was required first. After the update the recorder recognised the new cables normally.

Safety

Before returning the recorder to clinical use, inspect the leadset and patient cable for damage and run the manufacturer's electrical safety test for patient-applied parts, including leakage current.

Lesson: New accessories can require the current software or firmware version, so check version compatibility before blaming the cables, the connectors or the recorder itself.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.