Academy / Field fault notes / Interacoustics AD226
Interacoustics · AD226

Interacoustics AD226: headphone sound drops out when the cable is pressed

What happened

Sound in the audiometer headphones disappeared when pressing on the cable at one particular spot; releasing the pressure brought the audio back.

What was checked

Cause

In this case the cause was a fatigue break of the conductors in the cable at the connector, from repeated bending — a typical strain-relief failure. Pressing the cable at that spot opened the break and cut the audio path.

What was done

The headset cable proved to be detachable, secured with a small hex screw, so the engineers replaced the complete cable with an aftermarket equivalent rather than patching the break. The original tips were retained and insulated with heat shrink and tape, since the rubber boot could not be taken apart cleanly. Sound was stable after the replacement, as reported in the case; after any headset cable change, the audiometric output levels should be re-verified.

Safety

Because the headset is a patient-connected circuit, complete the electrical safety check (protective earth continuity and leakage currents) before returning the audiometer to use, and have the output levels re-verified after any headset cable replacement.

Lesson: An intermittent fault that follows pressure on the cable is a conductor break at the connector until proven otherwise — start with the detachable cable and its strain relief, not with the instrument.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.