Iris iQ200: Startup positive control fails
What happened
On a urine sediment analyzer, the primary positive control kept failing during quality control. No instrument error codes were reported in the case.
What was checked
- The engineers started from the reagents rather than the hardware: check the diluent for expiration date, how long the bottle has been open, and storage conditions.
- Verify that the diluent in use is suitable for this analyzer type; in the case, colleagues considered a compatible branded alternative acceptable.
- Repeat the positive control with fresh diluent and well-mixed control material before suspecting optics or fluidics.
Cause
No definite cause was established in the case. The engineers considered the diluent the likely factor, since on this analyzer it critically affects results, and an aged or unsuitable diluent can fail the positive control.
The cause is probable and was not fully confirmed.
What was done
The recommended fix, with only a probable outcome, was to replace the diluent with a fresh one, or with a compatible branded alternative intended for this type of urine analyzer. No confirmed result of this replacement was reported in the case.
Safety
Treat diluent, control material and sample waste as potentially infectious biological material: wear gloves and dispose of them per biohazard procedures.
Lesson: When quality control fails without error codes, check reagent freshness and suitability before opening the instrument.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.