Кардиотехника · 07-АД-1

Kardiotekhnika 07-AD-1: cuff will not inflate to set pressure

What happened

The unit could not build pressure in the cuff up to the set values. The cuff was new, and no obvious external leaks were noticed.

What was checked

Cause

The engineers considered the pump the most likely cause: reported in the case as a known wear item on this model that fails often, so a weak pump — or its sticking valves — cannot reach the set pressure even with a sound cuff and sealed circuit. The diagnosis was probable, not confirmed by further analysis.

The cause is probable and was not fully confirmed.

What was done

The fix in the case was pump replacement; per the report, the pump had already been changed more than once and is widely available as a spare part. After replacement, re-check the whole pneumatic circuit for leaks and confirm the relief valve operates. The outcome was reported as probable.

Safety

Before returning the unit to use, verify the overpressure relief function and pressure accuracy against a calibrated reference gauge with the cuff connected only to instruments — never trial the inflated cuff on a person.

Lesson: When a cuff will not inflate, work through the entire pneumatic path — cuff, connections, internal leaks, relief valve, pump — and treat the pump as a consumable, not a lifetime part.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.