Kasimov dry-heat sterilizer: slow warm-up and sterilization timer restarting

What happened

As reported in the case, the unit took more than 30 minutes to reach the sterilization regime, and whenever the temperature stepped 1 °C outside the 180 ± 1 °C window (182 °C was recorded), the sterilization countdown dropped to zero and started over, so a full cycle took far longer than expected.

What was checked

Cause

The manufacturer confirmed a defective batch: a firmware behaviour in which any 1 °C temperature deviation was treated as a failed hold, resetting the sterilization timer to zero — this also explains the very long time to reach a completed cycle.

What was done

The factory classified it as a known batch defect and asked for the unit to be sent back for reflashing as a paid manufacturer service; the reflash is a procedure from the manufacturer's service documentation, done by authorised personnel at the plant. After the update the behaviour was reported resolved.

Safety

Never release a load from an interrupted or restarted cycle as sterile — discard or reprocess it, and after any control or firmware work run a full validated cycle and complete electrical safety testing before returning the sterilizer to use.

Lesson: A cycle timer that silently restarts on a small deviation can stretch turnaround far beyond expectation, so reproduce and document the reset behaviour and get the manufacturer to confirm a batch or firmware defect before scheduling the repair.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.