Kasimov VKa-75: slow water fill into the steam generator
What happened
Between cycles, the steam generator refilled very slowly through the distilled-water line. The unit has no manual fill mode — only automatic programs — so every next start waited on the slow fill.
What was checked
- Walked the fill line from the distilled-water connection to the steam generator, noting which valves stand in the path and what holds each one closed.
- The engineers found that the generator outlet is held shut by a solenoid valve between cycles, so the trapped air had nowhere to escape and throttled the incoming water.
- Ruled out the valve and the water supply themselves — both worked as designed in this case.
Cause
The manufacturer's fill scheme relies on a solenoid valve that keeps the steam generator outlet closed between cycles. With the outlet sealed, the air inside acted as a lock, and the problem was venting rather than a failed part.
What was done
As a workaround, with the unit idle and depressurized, the engineer briefly freed the quick-disconnect tube running from the steam generator to the pressure gauge; once the air could escape, the generator filled far faster, and the tube was reconnected before the next cycle. The case also mentions a tidier long-term option: feeding the generator from a distilled-water tank through a small pump with an adjustable bypass. Treat the tube removal as a manual workaround, not a repair.
Safety
Never loosen any fitting on the steam generator until you have confirmed it is fully depressurized and cooled — trapped hot water or steam from a quick-release can scald. After any line has been opened, check it for leaks and confirm normal operation before returning the sterilizer to use.
Lesson: When a closed vessel fills slowly, look for the trapped air and give it a vent path before blaming the valve, the pump or the water supply.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.