Касимовский завод · ВК-75

Kasimovsky VK-75: steam generator repair and pressure-vessel paperwork

What happened

The sterilizer's steam generator was no longer fit for use and needed welding repair or replacement, and the work had to be properly documented as an intervention on a pressure vessel.

What was checked

Cause

No root cause was reported in the case; the reported problem was simply a steam generator unfit for further use, so the practical questions were the repair method and the pressure-vessel documentation.

What was done

Reported in the case: the steam generator was repaired on site by argon (TIG) welding, followed by a hydraulic pressure test. The work was closed out with a technical-condition report and a stamped entry in the repair-and-test log. Buying a new steam generator from the manufacturer was considered as an alternative. The outcome was reported as solved.

Safety

Treat the steam generator as a pressure vessel: isolate it, vent and drain it fully and let it cool before opening or welding, and prove the repair with a hydraulic test at the documented test pressure before the sterilizer returns to use.

Lesson: On pressure-vessel parts such as steam generators, a sound weld is only half the job — the hydraulic test and the properly stamped repair records are what allow the unit to legally return to service.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.