KLS Martin MAXIUM: recurring fault message on display
What happened
The electrosurgical unit kept showing a fault message (reported in the case as 50% on the display). Diagnostics pointed to the power / real-time clock subsystem, and after the internal battery was replaced the fault did not return.
What was checked
- Diagnostics were run (reported in the case: at the service centre); the engineers found the internal 3 V real-time clock battery discharged.
- When this kind of non-specific fault appears, check the internal clock battery first before blaming the main electronics.
Cause
A discharged 3 V coin-cell clock battery (CR2032 in this case). The engineers reported seeing this twice on such units; it is an observed pattern in their experience, not a general rule.
What was done
The internal coin cell was replaced with a new 3 V CR2032, and the fault cleared. Replacement means opening the enclosure of a mains-powered unit, so it should follow the manufacturer's service documentation.
Safety
Disconnect the unit from mains before opening the enclosure and, before returning it to clinical use, complete the electrical safety test (protective earth, leakage currents) and a functional check of the RF output per the manufacturer's documentation.
Lesson: A vague, recurring fault message can come from something as cheap as a dead clock battery — check the simple, low-cost parts before suspecting the main electronics.