KLS Martin · MAXIUM

KLS Martin MAXIUM: recurring fault message on display

What happened

The electrosurgical unit kept showing a fault message (reported in the case as 50% on the display). Diagnostics pointed to the power / real-time clock subsystem, and after the internal battery was replaced the fault did not return.

What was checked

Cause

A discharged 3 V coin-cell clock battery (CR2032 in this case). The engineers reported seeing this twice on such units; it is an observed pattern in their experience, not a general rule.

What was done

The internal coin cell was replaced with a new 3 V CR2032, and the fault cleared. Replacement means opening the enclosure of a mains-powered unit, so it should follow the manufacturer's service documentation.

Safety

Disconnect the unit from mains before opening the enclosure and, before returning it to clinical use, complete the electrical safety test (protective earth, leakage currents) and a functional check of the RF output per the manufacturer's documentation.

Lesson: A vague, recurring fault message can come from something as cheap as a dead clock battery — check the simple, low-cost parts before suspecting the main electronics.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.