Kodak · 2200

Kodak 2200: Intermittent beeping at power-on, relays do not engage

What happened

On power-up the unit emitted an intermittent beep and the relays did not engage; the power toggle switch had also sparked initially. Visual inspection showed no burn marks.

What was checked

Cause

The engineers found an open circuit in the monoblock: the lead wire of a ceramic high-value resistor, 200 MΩ as reported in the case, was broken. The break left no visible trace, which is why inspection showed nothing and the start-up and relay-enable sequence kept failing.

What was done

The faulty resistor was located and the break repaired by fitting a replacement ceramic high-value resistor (200 MΩ, as reported in the case). After the repair the unit powered up normally and the fault was solved.

Safety

Before probing the supply area, isolate the unit from mains and verify that storage capacitors are fully discharged — high-value ceramic resistors like this one often sit in high-voltage bleed circuits — and run an electrical safety test before returning the unit to service.

Lesson: A board can look perfectly clean and still hide an open high-value resistor that blocks the whole start-up sequence, so measure the chain with a meter instead of trusting the eye.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →