Kodak · DryView 5700

Kodak DryView 5700: low laser power aborts printing

What happened

Two DryView 5700 imagers in this case kept aborting jobs with error 607 — insufficient laser power. The fault was persistent, not a one-off glitch.

What was checked

Cause

The engineers found a degraded laser diode in the optical unit: the diode had lost output, and no cleaning or adjustment brought the power back. A batch of these printers was reported to have diodes that aged quickly.

What was done

No workaround was possible. The fix was replacing the laser optical unit, together with the post-replacement alignment and calibration — a procedure from the manufacturer's service documentation, done by authorised personnel. The fault cleared after the replacement.

Safety

Laser imager optics mean a live laser even with the cover off: work on a de-energised unit, never look into the beam path and never defeat interlocks.

Lesson: When laser power sits below spec after cleaning and a power cycle, treat the diode as spent and plan an optical unit replacement instead of chasing calibration.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.