Konica Minolta · DryPro Sigma

Konica Minolta DryPro Sigma: startup fault after first network setup

What happened

During commissioning, right after setting the IP address and rebooting, the printer came up with error 935 (as reported in the case). The case also notes errors 150/175 when the unit is rebooted with the top cover open.

What was checked

Cause

As the engineers found in this case, the unit rebooted after configuration with the top cover open. The interrupted startup prevented background finalisation tasks from completing, leaving the system software in a corrupted state that produced error 935 at boot.

What was done

The unit was recovered by reinstalling the system software from a verified image — a procedure from the manufacturer's service documentation, done by authorised personnel. After the recovery the printer booted normally and the fault was cleared.

Safety

Before opening any covers or working inside the unit, switch it off and isolate it from the mains, and after any software recovery run the electrical safety test before returning the printer to clinical use.

Lesson: Never interrupt a first boot or a post-configuration reboot: keep covers closed and let all background routines finish, because a broken finalisation can corrupt the system software.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.