Academy / Field fault notes / Konica Minolta Regius 110 + ImagePilot
Konica Minolta · Regius 110 + ImagePilot

Konica Minolta Regius 110: film not feeding from cassette

What happened

On a Regius 110 working with an ImagePilot console, the unit would not pull the imaging plate (film) out of the cassette — the feed did not start. No error code was reported in the case.

What was checked

Cause

The engineers found the cause in the film feed mechanism itself: a screw in the transport drive was defective, so the mechanism could not pull the plate out of the cassette. This is one reported case, not a general rule.

What was done

The faulty screw was replaced — in this case a matching screw taken from the side cover served as the donor — and normal feeding was restored; the case is reported as solved. The improvised screw should later be swapped for the correct part of the proper specification.

Safety

Isolate the unit from the mains and let all moving parts stop before removing covers — transport rollers and drive points can pinch — and verify a full, correct feed cycle before the unit returns to clinical use.

Lesson: When a feed or transport fault appears, rule out the simple mechanical causes — fasteners, rollers, linkages — before blaming drive electronics or software.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.