КРТ (KRT) · ОКО со столом Opera

KRT OKO (Opera table): protection trips after start-up, console keys keep flashing

What happened

On power-up the unit completes initialization and then the protection trips. On the control console all focus keys, the mode-switch key and an unlabelled 'hidden' key flash continuously. No errors are logged anywhere — not in the generator, not in teleservice, not in the software logs.

What was checked

Cause

The engineers found a loss of communication on the CAN bus between the control nodes. The built-in diagnostics of the generator and the application software simply do not flag this type of failure, so it is invisible to all the standard error stores — the only trace was entries in the Windows system event logs.

What was done

The engineers traced the CAN bus and repaired the fault in the line. After the repair communication was stable, the unit passed power-up initialization normally and the protection no longer tripped. The case is reported as solved.

Safety

Isolate the unit from the mains before working on bus wiring, and after any repair to the control system confirm that the protection circuits and exposure interlocks still trip correctly, with an electrical safety test before return to use.

Lesson: A clean fault log does not mean a clean system: on this control architecture a CAN bus failure can go completely unreported, so check the Windows system event logs for communication losses before you start replacing sensors or boards.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.