Early access: courses are being reviewed by field engineers
Service EngineerAcademy
LF · 664А

LF 664A: no cooling from the water loop

What happened

A surgical laser came in with cooling completely dead: the unit could not be run because the water loop was not removing heat.

What was checked

Cause

Reported in the case: the unit had been transported, the pump took in air and stopped moving water, and stale coolant had left sediment and sludge in the tubes — the loop simply could not carry heat any more.

What was done

No parts were fitted: the engineers flushed the tubing, replaced the water in the loop and bled the air out of the circuit — massaging the tubes by hand helps move the bubbles out of the pump. Cooling returned to normal.

Safety

Never fire the laser or leave it in a ready state without confirmed coolant circulation — an overheated emitter is a burn-through, fire and beam-hazard risk — and verify cooling plus an electrical safety test before returning the unit to use.

Lesson: After any transport, prime the cooling pump and refresh the water before power-up — trapped air and stale coolant are a common and cheap-to-fix reason a cooling loop dies.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes