LF 664A: no cooling from the water loop
What happened
A surgical laser came in with cooling completely dead: the unit could not be run because the water loop was not removing heat.
What was checked
- Checked the pump for an air lock — the engineers found it had lost its prime.
- Looked at the loop condition: the water was murky and sediment had settled in the tubing.
Cause
Reported in the case: the unit had been transported, the pump took in air and stopped moving water, and stale coolant had left sediment and sludge in the tubes — the loop simply could not carry heat any more.
What was done
No parts were fitted: the engineers flushed the tubing, replaced the water in the loop and bled the air out of the circuit — massaging the tubes by hand helps move the bubbles out of the pump. Cooling returned to normal.
Safety
Never fire the laser or leave it in a ready state without confirmed coolant circulation — an overheated emitter is a burn-through, fire and beam-hazard risk — and verify cooling plus an electrical safety test before returning the unit to use.
Lesson: After any transport, prime the cooling pump and refresh the water before power-up — trapped air and stale coolant are a common and cheap-to-fix reason a cooling loop dies.