Maquet Flow-i: gas analyser test fails on low oxygen concentration

What happened

On this Flow-i the gas analyser self-test kept failing, and the failure reason was a low oxygen concentration reading. The case also raised practical points about battery replacement on this model.

What was checked

Cause

The oxygen concentrator feeding the machine was delivering an O2 concentration below what the Flow-i accepts. The case reported this model as demanding about inlet oxygen concentration, so the analyser test failed on supply quality — the analyser itself was healthy.

What was done

Bring the oxygen supply up to the concentration the machine requires (or replace the source with one that meets it), then rerun the self-test and the full pre-use check. Battery replacement on this model is a procedure from the manufacturer's service documentation, done by authorised personnel; the engineers in this case recommended replacing the battery roughly every five years.

Safety

Before returning the machine to use, verify the oxygen analyser against a gas mixture of known concentration and complete the manufacturer's full pre-use check — a machine that delivers wrong oxygen readings to a patient is a direct clinical hazard.

Lesson: When a gas analyser test flags low oxygen, verify the oxygen supply upstream first — the test may be honest and the fault can sit in the source gas, not in the machine.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.